SPECIFICATIONS:
Model: B-2000
POWER
Power Requirements: AC220V 10%, 50Hz 2%
Power Input: ≤450VA
Refer to IEC60601-1, this equipment belongs to Class I, Type B. Its operation is Continuous.
ALARM
Power Alarm
Power Failure: Activates if primary power to the Incubator
Fails or the power wire is accidentally
Disconnected from the wall receptacle.
Fan Alarm
Fan Failure: Activates if Fan motor malfunctions. Sensor Alarm
Air Sensor Failure: Air Sensor malfunction, refer unit to service.
Sensor Out of Position: Activated when Sensor is not in correct position To perform calibration or continue operation.
Skin Sensor Disconnected: Activated during Skin Control Mode when Skin Sensor is disconnected from Temperature Controller. Skin Sensor Failure: Activated when Skin Sensor malfunctions.
Deviation Alarm High Air Temperature: Activated if displayed Air Temperature is 1.5 centigrade above set point
Low Air Temperature: Activated if displayed Air Temperature 2.5centigrade below set point
Skin Temperature: Activated if displayed Skin Temperature is 1.0 centigrade above set point
Low Skin Temperature: Activated if displayed Skin Temperature is 1.0Over Temperature Alarm
Over Air Temperature: Activated when displayed Air TemperatureWas approach 38centigrade for set temperature ≤37centigrade, Or approach 40centigrade for set temperature >37centigrade in Air Mode.
Over Skin Temperature: Activated when displayed Skin Temperature was approach 38centigrade for set temperature ≤37centigrade, Or approach 39centigrade for set temperature >37centigrade in Skin Mode. System Alarm
System Failure: Activated if Temperature Control System malfunctions.
TEMPERATURE CONTROLRANGEAND CORRELATIVE SPECIFICATIONS
Air Temperature: 20centigrade ~37centigrade, for set temperature ≤37centigrade
37centigrade ~39centigrade, for set temperature >37centigrade
Skin Temperature: 20centigrade ~37centigrade, for set temperature ≤37centigrade
37centigrade~38centigrade, for set temperature >37centigrade
Temperature Rise Time: ≤30min
Temperature Variability: 0.5centigrade
Temperature Uniformity: ≤ 1 centigrade
Skin Sensor Precision: ≤0.3centigrade
ENVIRONMENT REQUIREMENT
Environment Temperature
Operation Temperature Range+20centigrade ~+30centigrade
Store Temperature Range -40centigrade ~+70centigrade
Environment Humidity
Operation Humidity Range: 30%~75%
Store Humidity Range: ≤80%
Atmospheric Pressure Range: 500hPa~1060hPa
OTHER SPECIFICATION
Noise Level Within Hood Environment ≤55dB(Environment noise below 45dB Carbon Dioxide (CO2) Level within the Hood Less than 0.5% when a mixture of CO2
In air is delivered at 750ml/min
At point 10cm above the center of the mattress.
Air Velocit y Over Mattress less than: 0.35m/s
Infant Bed Max Load: 10kg
Transfusion Shelf Max Load: 2kg
Shelf Max Load:
Cabinet Max Load: 6kg