  • TKR-200C ventilator for beasties

    TKR-200C ventilator for beasties that is device according to the principle of open jet ventilator can ventilate with not only high frequency and low volume of humidifying but also regular frequency and great volume.

    A. When the ventilation technique with open high frequency and low humidify volume is adopted, the airway pressure is low with little effect of oxygen compound is good and the oxygen partial pressure can be increased rapidly. While the one with regular frequency and great humidify volume is adopted, CO2 can be ejected easily.

    B. Convenient connecting, it can be connected through the nasal and endotracheal tube.

    C. The ventilation mode with high frequency and low humidify volume will not easily produce pressure hurt.

    D. The effect of oxygen dispersion is so good that the tested beastie can stand for a long time.

    E. Adjustable oxygen concentration, the ventilation mode with continuous oxygen stream and good humidifying effect can ensure beastie to respire independently.

    F. Be fit for manual ventilation test for monkey, rabbit, dog etc.

    G. Configured with multiform ventilation conduit for beastles or connected with gas source, in addition.



    Specifications:

     Frequency:8~150 breaths/min

     I/E ratio:1:1~1:3 and inverse ratio

     Driving Pressure :0~0.08MPa

     Gas source: High-pressure oxygen source or air compredsor. The input pressure is not more than 0.16MPa

     High-pressure gas source:0.16MPa

     Output Limit of reductor: not less than 0.16MPa

     Ventilation Mode:HFJV(High Frequency Jet Ventilation);CPAP(Continuous positive airway pressure),Manual Control Mode.
     


  • TKR-300H computerized high-frequency jet ventilator

    TKR-300H computerized high-frequency jet ventilator is a new kind of opening high frequency jet ventilator controlled by singlechip, it adopts opening ventilation mode with high frequency and low humidifying volume so that the airway pressure is low with little effect on blood stream recycling to heart and it helps to increase the blood flux from the heart and to promote the oxygen combination. The operation does not need to keep the pace with respiration, the device can supply oxygen accessorily and controlled ventilation as well. When ventilating, the suction in airway can be adjusted at the same time.

    A. It is widely used for the respiratory management, treatment and first-aid resuscitation of animal such as cattle, horse, tiger, elephant, panda and so on.

    B. Multiform switching mode for jetting oxygen through the nose, the device can improve oxygen density rapidly and has an obvious effect on the treatment for all kinds of hypopnea.

    C. The device can ventilate through the endotrachea tube and its operation does not need to keep synchronous with respiration. It can be used for tracheotomy in an emergency.



    Specifications:

     Frequency:10~150 breaths/min

     I/E ratio:1:1~1:3 and inverse ratio

     Driving Pressure :0~0.20MPa

     Process ventilation: T=10min,

    Pro I :f(150):f(35)=7:3,

    Pro II :f(35):f(150)=7:3,

     Alarm:Alarm with light and sound for no power, and last 6 seconds if no oxygen

     CPAP:not less than 0.05-1.0KPa 

     PEEP:not less than 0.05-0.5KPa

     Oxygen Concentration Adjustable:40%-90%

     Jetting volume range:0-4000ML/time

     Pressure Regulator Limit:0.30MPa

     Ventilation Mode:HFJV(High Frequency Jet Ventilation);CPAP(Continuous positive airway pressure),Pro I and Pro II ventilation,HFJV+SIGH ventilation; HFJV+PEEP ventilation and so on.
     


  • TKR-400T computerized multi-function high-frequency ventilator

    Feature:
    A. HF JV(High-frequency jet ventilation) with pressure support, which can supply the higher inspiration pressure so that the patients can inspire sufficient oxygen with a wet finger.
    B. Opening airway. The ventilation can be made with high frequency and little tidal volume and low airway pressure, which is good to increase the blood oxygen density.
    C. Ventilation mode through nasal vestibule without effect which can avoid from tracheotomy and endotracheal tube so as to decrease patient's pain and the effectiveness of oxygen feeding is very good, The patient can talk, drink and eat, when ventilating.
    D. Continuous positive airway pressure with adjustable oxygen density plus effective humidifier can show the feature of CPAP better.
    E. The multi-functional humidifier adaptor can be connected with any type of humidifier by trachea intubation of national or international standard.
    F. The pressure humidifier has knob to manipulate the air volume and has the delicacy alarm with sound and light for no power or gas.

    Specifications:
     Frequency:10~150 breaths/min
     I/E ratio:1 : 1~ 1 : 4 and inverse ratio
     Driving Pressure :0~0.15MPa 
     Process Ventilation:T=10min,
    ProI: f(150):f(35)=7:3
    ProII: f(35):f(150)=7:3
     SIGH:3,6,9breaths/min
     Alarm:Alarm with light and sound for no power and gas, lasting 6 seconds if no oxygen
     CPAP:not less than 1.0kPa
     PEEP:not less than 0.5kPa
     Oxygen Concentration Adjustable:40%-90% displayable
     Jetting volume range:O~1400ml/time
     Pressure Regulator Limit:0.03MPa
     Ventilation Mode:HFJV(High Frequency Jet Ventilation);CPAP(Continuous positive airway pressure);ProI and ProII ventilation; HFJV+PEEP ventilation and so on


  • TKR-300CG high-frequency jet ventilator

    Feature:
    A. The device adopts impulse jet oxygen feeding technique with great flow rate of oxygen feeding in a shot while, when feeding oxygen, it adopts change mode major in reflux and many kinds of gas flow mechanism such as convection, laminar flow, sheet flow and so on concurrently, which is good to increase partial pressure of oxygen rapidly, At the same time fluctuation of thorax caused by increasing oxygen flow into airway is good to let out CO2.
    B. Jet oxygen feeding is supported by pressure when inspiration and secure certain PEEP effect when feeding oxygen, So as to reach good effectiveness of oxygen feeding.
    C. Accumulator jar, it can be used while being charged and can last for four hours after having been charged.
    D. Alarm system with sound and light when lack of electric quantity or airway pressure.
    E. 300CG chooses digital display of import pressure sensor.
    F. 300CG has three kinds of airway limited pressure alarm for adult, child and infant.
    G. The device disposes pressure humidifier which has good humidigying effectiveness.

    Specifications:
    Frequency: 10~150 breaths/min
     Power supply:AC220V±22V 50HZ±1HZ
     I/E ratio:1:1,1:1.5~ 1:3 and 3:1
     Driving Pressure :0~0.15MPa 
     SIGH:6breaths/min, I/E=1:4
     Pressure Regulator Limit:0.30MPa
     CPAP:not less than 1KPa 
     Oxygen Concentration Adjustable:40%-90%
     Process Ventilation:T=10min,Infinite recursion
    ProI: f(150):f(35)=7:3
    ProII: f(35):f(150)=7:3
     Alarm:Alarm with light and sound for no gas or no power or airway pressure is over
     Inner Battery:4Ah 12V
     Ventilation Mode:HFJV(High Frequency Jet Ventilation);CPAP(Continuous positive airway pressure),HFJV+SIGH(Sigh); Process ventilation I and Process ventilation II, Manual Control Mode.
     


  • TKR-400A computerized high-frequency ventilator for infant

    Feature:

    It is a ventilation device for infant with novel structure and good performance, and chooses opening ventilation mode with high frequency and little tidle volume, so the airway pressure is low.

    A. The tidal air can letout with constant pressure, constant current or accurately adjustable flow.

    B. Opening ventilation, opening intubation makes it not need casing bag, little resistance as expiration is suited for the ventilation management of infant.

    C. Evident effectiveness of treating ARDS can be shown with inserting small tube and jetting oxygen through nasal vestibule, which is secure and doesn't cause pressure wound easily.

    D. PEEP and humidifier volume can be adjusted accurately.

    E. Alarm system with sound and light when no oxygen, poor of electric quantity or low airway positive pressure.

    F. Heated humidifier can be used for good effectiveness of humidifying.

    G. Accumulator jar, it can be used while being charged and can last for more than four hours.



    Specifications:

     Frequency:10~150 breaths/min

     I/E ratio:1:1,1:1.5~ 1:3 and inverse ratio

     Driving Pressure :0~60KPa 

     Pressure Atomizer Rate:>=6ml/h

     Process Ventilation:T=10min,

    ProI: f(150):f(35)=7:3

    ProII: f(35):f(150)=7:3

     Storage Battery:4Ah, Automatic charge when 12V

     SIGH:3breaths/min

     Alarm:Alarm when battery voltage is lower than 11V, alarm lasts 6 seconds if no oxygen, Alarm with audible signals and flashing light when the airway pressure is over 0.8KPa

     CPAP:0~0.5KPa 

     PEEP:0~0.20KPa

     Oxygen Concentration Adjustable:21%-90%

     Jetting volume range:O~600ml/time

     Ventilation Mode:HFJV(High Frequency Jet Ventilation);CPAP(Continuous positive airway pressure);HFJV+PEEP(Positive End-Expiratory Pressure), HFJV+SIGH(Sigh), Process ventilation I and Process ventilation II, Manual Control Mode.

     


  • TKR-300JII computerized high-frequency first-aid ventilator

    Feature:

    It is widely used for clinical salvaging and revival also is a dedicated device used for respiration salvaging on emergency ambulance.

    A. An opening system is designed to avoid leakage of airway, When ventilating, the intubation may have no casing bag. Little tube may be used for ventilation when resistance of expiration is small, so that mucose won’t be hurt easily.

    B. Airway is open during wentilarion, suction and ventilation can be done synchronously, so good effectiveness of oxygen combination can be shown without inspiration and expiration synchronizing.

    C. Lower airway pressure is good to increase blood discharge and decrease encephal pressure.

    D. Suction device, its driving negative pressure is adjustable.

    E. Accumulator jar, it can be used continuously for not less than four hours after being charged.

    F. The system disposes completely with laryngoscope, inlet intubation, catheter of tracheotomy, nasal prongs tube, and air tube etc.



    Specifications:

     Frequency:10~150 breaths/min

     I/E ratio:1:1,1:1.5~1:3 and inverse ratio

     SIGH:6breaths/min,

     CPAP:≥1.5KPa 

     Suction negative pressure:-0.05~0MPa

     Driving Pressure :0~0.15MPa

     Constant volume ventilation: T=10min, I/E=1:1.5

    Adults:f(100):f(20)=7:3, tidal volume=550ml/time, Driving pressure=0.06MPa

    Children:f(100):f(20)=7:3, tidal volume=350ml/time, Driving pressure=0.04MPa

     Storage battery:4Ah, Automatic charge when 12V

     Most jetting volume:≥1400ML/time

     Oxygen Concentration Adjustable:40%-90%

     Alarm:Alarm when battery pressure is lower than 11V; Alarm when the oxygen pressure is lower than 0.01MPa; Alarm with sound and light when no gas

     Ventilation Mode:HFJV(High Frequency Jet Ventilation);CPAP(Continuous positive airway pressure),HFJV+SIGH; Constant volume ventilation.

     