PM-9000E+ Fetal Maternal Monitor
Fetal Heart Rate
l Probe: nine-unit, wide range, pulse wave, high sensitivity
l Signal Strength: < 1 mw/cm2
l Frequency: 1.0 MHz
l Signal processing: DSP processor
l Max Sound Power: 1.5 watt
TOCO Pressure
Measure Range: 20 – 100
Pressure ultrasonic probe output
In accordance with international practice IEC1157 1992, the peak negative pressure less than 1 Map (Map). Beam intensity is less than 20 mW / cm 2 (mW/cm2). Space instantaneous peak intensity average density is less than 100 mW / cm 2 (mW/cm2). The type of sound intensity less than 1mW/cm2
Coupling agent (GEL)
Viscosity of water-based compounds no stimulates to the skin, no allergies. Chemical stability, inhibit bacteria.
ECG measurement:
Leads choices: All, I, II, III, aVF, aVR, aVL. V1--6 chest leads showed
Gains options: × 1, × 2, × 4, and × 0.5 four trenches
Frequency: 0.5 Hz - 100 Hz
Common-mode signal inhibition: More than 70 dB
Scan rate: 12.5 mm / S, 25 mm / S, 50 mm / S standard three stalls.
HR: ± 5 bpm
Precision calibration signal: 1mv(peak, ± 3% accuracy)
ST Detection: measurement range (-1.0mV----+1.0mV)
Using Standard: meet the ANSI / AMI EC13 -1992 standard
Response time: 1s
Alarm: Sound and Light
Preset heart rate range: (Alarm over preset numerical)
Ceiling: 1-250 bpm adjustable
Threshold: 0-249 bpm adjustable
NIBP measurement
Technology: Oscillometric
Mode: manual / automatic
Measurement units: mmHg / kpa
Protection: Double security protection
A measurement time: less than 45 S
Scope: Pressure Cuffs 0-250 mmHg
Blood pressure measurement accuracy: ANSI /AAMI, SP10-1992 standard error and standard deviation of the average value of the initial
Inflatable cuffs: 180 mmHg (120 mmHg newborn children 70 mmHg)
Inflatable cuffs follow: original SBP: + 30 mmHg
Blood pressure measuring range:
Systolic blood pressure: 60-250 mmHg
Diastolic blood pressure: 40-220 mmHg
Pressure on the average: 45-235 mmHg
Response time: 1s
Alarm: Sound and Light
Preset range: report (Numerical exceed preset alarm)
Systolic blood pressure maximum: 1-255 mmHg adjustable
Threshold: 0-254 mmHg adjustable
Diastolic blood pressure ceiling: 1 -255 mmHg adjustable
Minimum standards: 0 -254 mmHg adjustable
Use standard: compliance with the ANSI / AAMI SP10-1992 standard
Spo2 measurement
Range: 0-100%
Accuracy: 1%
Resolution: 1%
Pulse frequency: 20-250 times / min
Response time: 1s
Alarm: Sound and Light
Preset range: report (Alarm over preset numerical)
Ceiling: 1% - 100% adjustable
Threshold: 0% - 99% adjustable
TEMP measurement:
Technology: thermostat temperature probe temperature
Measuring range: 20 ° C-45 ° C
Temperature resolution: 0.1%
Accuracy: ± 1 ° C
Response Time: 1s